三维超声心动图
三维超声心动图(3DE)是一种超声方法,可采集心脏的容积性动态图像,使心腔和瓣膜得以测量和显示,而非像常规超声心动图那样只能显示切面。由于容积直接从三维数据集测量,左心室(LV)容积和射血分数无需依赖限制二维成像的几何假设和缩短效应即可获得;指南文件将3DE对左心室容积和射血分数的评估描述为比二维超声心动图更准确、重复性更好,其准确性与心脏磁共振(CMR)相似。1 美国超声心动图学会(ASE)和欧洲心血管影像协会(EACVI)的心腔定量指南推荐,在具备足够专业经验的实验室中,对左心室和右心室采用3DE测量。2
| 关键事实 | 数值 |
|---|---|
| 换能器 | 矩阵阵列,约2,500–3,000个压电元件,而常规二维相控阵为128个3 • 4 |
| 采集模式 | 实时(单心动周期)3DE,或心电门控多心动周期窄子容积拼接,覆盖2–7个心动周期1 |
| 左心室全容积采集 | 一般在10秒内完成,于屏气时从心尖窗口采集1 |
| 左心室容积与CMR比较 | EDV r = 0.91,ESV r = 0.93,但在一项多中心验证中容积较CMR低26%和29%5 |
| 与CMR的合并偏倚(半自动软件) | EDV −39.3 mL,ESV −19.6 mL,EF −0.6%6 |
| 二尖瓣反流的三维缩流颈面积 | 截断值0.41 cm²:重度MR的敏感度97%,特异度82%7 |
工作原理
其物理基础是对容积进行电子扫描,而非探头的机械运动。当前的矩阵阵列换能器由近3,000个压电元件组成(相比之下,常规二维相控阵换能器为128个元件),经胸成像工作频率为2–4 MHz,经食管成像为5–7 MHz。3 据一项描述,2,500个晶片按五十行五十列排列,每个晶片独立激活、聚焦和偏转,以覆盖一个锥体形扫描容积。4
快速填充容积需要并行波束形成。要在16 cm深度以25容积/秒获得90° × 90°的全容积锥体,系统必须每秒接收100,000–200,000条波束,这要求每个发射脉冲有20至40条并行接收波束。3 图像生成随后经过四个步骤:数据采集、存储、处理(转换为体素并插值)和显示。4 空间分辨率呈各向异性:点扩散函数模糊在轴向约0.5 mm,侧向约2.5 mm,仰角方向约3 mm,因此胸骨旁成像(轴向波束方向)对结构的解析优于心尖视图。3
操作方法
经胸工作流程从二维图像优化开始,经过采集模式选择(窄角与宽角/全容积、单心动周期与多心动周期、三维放大、三维彩色),到渲染和最终显示,以及分析。8 心电门控要求R波清晰可见,这对触发模式成像至关重要,而较新的“一体式”换能器可用单一探头采集二维和三维图像。8 总增益通常需要比二维成像增加50%以上(约55–60个单位),而较大的采集容积会降低帧率。8
模式选择取决于目标。实时窄扇区模式显示30° × 60°的锥体容积;门控全容积模式扇区最大,时间分辨率高于30 Hz,适用于二尖瓣或主动脉根部。1 全容积是心室和心房容积及射血分数的首选模式,而实时三维放大适用于瓣膜解剖、分流和经导管引导。8 采集前裁剪可获得更好的分辨率,而采集后裁剪宽数据集则保留更多诊断信息1;裁剪可使用固定断层平面、自由平面或由二维横截面引导的平面。7 对于经食管检查,存在五种模式(同步二维多平面、单心动周期实时、放大、心电门控多心动周期合并最多六个子容积,以及三维彩色多普勒);多心动周期模式需要规整心律和屏气,因此对于不规则心律和操作引导,单心动周期放大模式更受青睐。7
起源
1974年,唐·L·德克尔、罗伯特·L·皮齐亚利和尤金·董在《计算机与生物医学研究》上发表了首例超声人体心脏三维重建,该系统通过记录探头位置将多个二维横截面组装成三维图像。9 向实时容积成像的转变伴随着奥拉夫·T·冯·拉姆和斯蒂芬·W·史密斯于1990年在《数字成像杂志》(现更名为《医学成像信息学杂志》)上描述的实时容积超声成像系统原型而到来。10 • 29 早期的稀疏阵列方法使用256个元件非同时激活,在单次心跳内生成60° × 60°的锥体容积,但无法显示实时渲染的三维图像,且图像质量差、换能器接触面大、缺乏便携性限制了其应用11;杜克大学的机器非常庞大,尽管采用并行处理,帧率也仅达到约8容积/秒。12 张全斌及其同事于2013年在《国际心脏病学杂志》上报道了经CMR验证的、用于右心室容积定量的单心动周期全容积采集。13 后来带有在线渲染三维显示的完全采样矩阵阵列使该方法更接近常规临床应用。11
变体
三维斑点追踪超声心动图在灰阶全容积数据集中追踪超声斑点,得出纵向、周向、径向和面积应变,以及旋转、扭转和扭矩;由于所有分量均来自一个三维数据集(根据采集模式可能由多个门控心动周期组装而成),分析所需时间约为二维斑点追踪的三分之一。14 全容积三维斑点追踪采集在呼气末屏气时进行,覆盖最多六个心电门控心动周期,要求容积率高于心率的40%或超过25帧/秒。14 一种用于从三维数据集分析右心室容积的专用斑点追踪算法由德妮莎·穆拉鲁及其同事于2015年在《欧洲心脏杂志-心血管影像》上经CMR验证。15
对于二尖瓣,离线动态三维模型使用专有软件构建:飞利浦QLAB MVQ、TomTec 4D MV Assessment和西门子eSie Valves,其中MVQ/MVN(二尖瓣定量/导航)分析在专用软件包中进行。4 • 7 商用三维/四维经食管超声探头包括5–7 MHz的X7-2t(飞利浦)和3–8 MHz的6VT-D(通用电气)。4 自动化已从半自动发展到基于机器学习的右心室定量,分别由达维德·杰诺韦塞及其同事于2019年在《美国超声心动图学会杂志》16和大谷京子及其同事于2019年在《欧洲心脏杂志-心血管影像》17上经CMR验证,多视图三维融合处理由泰勒·兰姆及其同事于2021年在《医学与生物学超声》上描述。18 基于视频的AI用于逐搏评估心脏功能由欧阳德及其同事于2020年在《自然》上报道19,用于三维超声心动图左心室分割的机器学习数据集MITEA(使用CMR衍生标签)由赵黛比及其同事于2023年在《心血管医学前沿》上发表。20
应用
心电门控全容积多心动周期模式被描述为实时3DE对二尖瓣进行解剖和功能表征的金标准21,三维经食管超声(3D-TEE)在识别退行性病因(如Barlow病和纤维弹性缺陷)方面比二维经胸超声(2D-TTE)更准确。21 三维经食管超声以正面“外科医生视图”显示瓣膜,有助于预测修复的复杂性和成功率22;在112例接受二尖瓣修复的患者中,佩皮及其同事发现三维经食管超声识别所有二尖瓣病变的准确率为95.6%,而三维经胸超声为90%,二维经食管超声为87%,二维经胸超声为77%。22 反流严重程度用三维缩流颈面积量化(截断值0.41 cm²:重度MR的敏感度97%,特异度82%)。7
其他成熟用途包括房间隔缺损尺寸测量(复杂ASD中三维经食管超声面积2.8 ± 1.3 cm²,而二维经食管超声为1.7 ± 1.4 cm²)22、左心耳封堵器尺寸选择(三维经食管超声的器械尺寸一致加权Kappa为0.62,而二维经食管超声为0.28)22、经导管主动脉瓣置换术(TAVR)瓣环尺寸测量(三维经食管超声根部测量与CT吻合良好,可在肾病患者中替代增强CT)2,以及先天性心脏病的心腔定量,其中一项纳入95项研究、3,055名受试者的荟萃分析发现3DE与MRI比较结果良好,但偏向于低估左心室舒张末期和收缩末期容积。23
以CMR为参考,一项多中心研究发现实时三维超声心动图(RT3DE)左心室容积高度相关(EDV r = 0.91;ESV r = 0.93),但低26%和29%,且偏倚与观察者经验呈负相关。5 从CMR参考中排除小梁和二尖瓣平面消除了模态间偏倚,表明将小梁计入心肌是主要误差来源。5 一项2023年荟萃分析发现RT3DE与CMR在LVEF、左心室质量、右心室收缩末期容积或RVEF方面无显著平均差异,但显著低估LVESV、LVEDV和RVEDV;在老年人和患病心脏中一致性较低。24 一项纳入38项研究、1,881名受试者的荟萃分析量化了软件效应:半自动软件的合并偏倚为EDV −39.3 mL、ESV −19.6 mL和EF −0.6%,而全自动软件对EDV和ESV的绝对偏倚较小(−14.5 mL和−6.3 mL),EF偏倚的绝对值略大(−1.1%对比−0.6%)。6 对于跨模态的射血分数,一项大型荟萃分析发现CT和3DE与CMR对LVEF的偏倚最低、一致性最好,而二维超声心动图的一致性界限最大(−13.3至12.1%,r = 0.660);对于RVEF,CT和3DE的偏倚低于5%,r > 0.75。25
局限性与替代方案
多心动周期拼接采集容易产生拼接伪影,即由心律失常、电灼和探头移动引起的子容积之间的分界线4,并且需要规整心律和长时间屏气。24 空间和时间分辨率仍不及二维超声心动图,分析可能耗时。24 增益设置很重要:增益过高会产生棕色斑点噪声,而信号丢失呈黑色,由增益不足、声影或衰减引起。4 在心律失常时,多心动周期采集不可行14,在婴儿中单心动周期采集可能缺乏高心率所需的时间分辨率,除非感兴趣区较小。23 准确性取决于操作者经验,如偏倚与经验的负相关关系所示。5
心脏结构和功能的自动测量如今在当前超声心动图机器上已很普遍,处于技术成熟度9级,尽管许多其他AI定量模型仍在开发中;基于AI的心脏定量软件已获得多项监管批准,例如美国食品药品监督管理局(FDA)于2025年9月批准Ligence Heart,2026年5月批准带Auto EF的uSONIQUE设备,2026年6月批准带PCWP附加组件的CorVista系统。26 对于基于AI的3DE,稳健性遵循层级:左心室容积最稳健,其次是射血分数,再次是左心房容积,后者最不稳健。27 训练与部署条件之间的域不匹配可能导致“无声漂移”,即隐藏在看似精确的轮廓背后的逐渐准确性损失27;专家指南建议经验不足的操作者不要仅依赖AI,在图像质量欠佳时应优先手动测量。27 2025年ASE/EACVI应变共识声明指出,由于软件变异性和结果不一致,三维应变仍处于开发阶段,不推荐用于临床。28 在硬件方面,厂商已开发出波束形成技术,如飞利浦nSIGHT、通用电气XDclear/cSound和西门子相干容积形成,彩色多普勒容积现在采集速度可达40容积/秒。3
References
- EAE/ASE Recommendations for Image Acquisition and Display Using Three-Dimensional Echocardiography (JASE, 2012)
- State-of-the-Art Review: 3-Dimensional Echocardiography, Latest Developments and Future Directions (JACC: Cardiovascular Imaging)
- Principles of Three-Dimensional Ultrasound (book chapter)
- Three-dimensional transesophageal echocardiography: Principles and clinical applications
- Real-Time 3-Dimensional Echocardiographic Quantification of Left Ventricular Volumes: Multicenter Study for Validation With Magnetic Resonance Imaging (JACC Cardiovasc Imaging)
- Accuracy of LV Volumes and EF by Contemporary 3DE with Semi- and Fully Automated Software: Meta-Analysis of 1,881 Subjects (JASE, 2019)
- Three-dimensional transoesophageal echocardiography: how to use and when to use, EACVI clinical consensus statement (Faletra et al., 2023)
- Three-Dimensional Echocardiography: Image Acquisition (Essential Echocardiography: Companion to Braunwald's Heart Disease, 2023)
- A system for ultrasonically imaging the human heart in three dimensions (Computers and Biomedical Research, 1974)
- Olaf T. von Ramm, Stephen W. Smith (1990). Real time volumetric ultrasound imaging system. Journal of Imaging Informatics in Medicine.
- Three-Dimensional Echocardiography: The Benefits of the Additional Dimension (JACC)
- A concise history of echocardiography: timeline, pioneers, and landmark publications
- Quan Bin Zhang and colleagues (2013). Feasibility of single-beat full-volume capture real-time three-dimensional echocardiography for quantification of right ventricular volume: Validation by cardiac magnetic resonance imaging. International Journal of Cardiology.
- Real-time three-dimensional speckle tracking echocardiography: technical and clinical aspects (Sorrentino et al.)
- Denisa Muraru and colleagues (2015). New speckle-tracking algorithm for right ventricular volume analysis from three-dimensional echocardiographic data sets: validation with cardiac magnetic resonance and comparison with the previous analysis tool. European Heart Journal - Cardiovascular Imaging.
- Davide Genovese and colleagues (2019). Machine Learning–Based Three-Dimensional Echocardiographic Quantification of Right Ventricular Size and Function: Validation Against Cardiac Magnetic Resonance. Journal of the American Society of Echocardiography.
- Kyoko Otani and colleagues (2019). Accuracy of fully automated right ventricular quantification software with 3D echocardiography: direct comparison with cardiac magnetic resonance and semi-automated quantification software. European Heart Journal - Cardiovascular Imaging.
- Tyler Lamb and colleagues (2021). Multi-View 3-D Fusion Echocardiography: Enhancing Clinical Feasibility with a Novel Processing Technique. Ultrasound in Medicine & Biology.
- David Ouyang and colleagues (2020). Video-based AI for beat-to-beat assessment of cardiac function. Nature.
- Debbie Zhao and colleagues (2023). MITEA: A dataset for machine learning segmentation of the left ventricle in 3D echocardiography using subject-specific labels from cardiac magnetic resonance imaging. Frontiers in Cardiovascular Medicine.
- Clinical Utility of Three-Dimensional Echocardiography in the Evaluation of Mitral Valve Disease: Tips and Tricks (J Clin Med, 2023)
- Clinical and perioperative applications of three-dimensional echocardiography
- Expert consensus for 3D echocardiography in congenital heart disease (J Am Soc Echocardiogr 2017;30:1-27)
- Validating real-time three-dimensional echocardiography against cardiac magnetic resonance, for the determination of ventricular mass, volume and ejection fraction: a meta-analysis (Clin Res Cardiol, 2023)
- Accuracy of cardiac CT, radionucleotide and invasive ventriculography, two- and three-dimensional echocardiography, and SPECT for LVEF and RVEF compared with cardiac MRI: a meta-analysis (Eur Heart J Cardiovasc Imaging, 2015)
- Artificial intelligence in echocardiography: a position statement from the British Society of Echocardiography (Echo Research & Practice, 2026)
- Artificial intelligence–powered three-dimensional echocardiography in clinical practice: the illusion of automation? (Eur Heart J Imaging Methods and Practice, 2026)
- Clinical Applications of Strain Echocardiography: A Clinical Consensus Statement From the ASE in Collaboration With the EACVI (2025)
- scholars.duke.edu
Initially written Sep 29, 2026 · Reviewed: Sep 30, 2026 · Edited: Sep 30, 2026