脾脏硬度测量(spleen stiffness measurement,SSM)是一种无创弹性成像技术,以千帕(kPa)为单位量化脾脏硬度,用于评估慢性肝病中的门静脉高压。由于脾静脉汇入门静脉,门静脉压力升高导致脾脏淤血,同时内脏高动力循环使脾脏硬度增加,因此脾脏变硬提示门静脉高压,特别是临床显著门静脉高压(clinically significant portal hypertension,CSPH),后者定义为肝静脉压力梯度(hepatic venous pressure gradient,HVPG)≥10 mmHg。1 SSM已通过瞬时弹性成像、点剪切波弹性成像、二维剪切波弹性成像和磁共振弹性成像(magnetic resonance elastography,MRE)得到验证,是门静脉压力的动态替代指标。1 它既反映脾脏的慢性结构改变——如脾脏血管结构变化和纤维化——也反映门静脉高压的动态血流动力学淤血。2 SSM用作肝脏硬度测量的辅助手段,在无创检测已改变cACLD患者CSPH诊断格局、而HVPG测量仍为有创且可及性有限的情况下,优化代偿期进展性慢性肝病(compensated advanced chronic liver disease,cACLD)的风险分层。3 • 4

关键事实详情
测量内容通过瞬时弹性成像(VCTE)、点SWE、二维SWE或MRE测量的脾脏硬度(kPa)1
生理学基础脾静脉汇入门静脉;淤血和内脏高动力循环使脾脏硬度增加5
参考标准CSPH = HVPG ≥10 mmHg(HVPG >5 mmHg定义为门静脉高压)1
Baveno VII截断值病毒性肝炎相关cACLD中,TE-SSM <21 kPa排除CSPH,>50 kPa确诊CSPH1
单截断值模型在Baveno VII标准基础上加入SSM ≤40 / >40 kPa,将不确定灰区从48%降至9%6
可行性专用100 Hz FibroScan模块总成功率达97.3%(1782/1832次测量)7
正常值健康成人SSM均值估计为18.35 kPa8

工作原理

所有弹性成像方法均基于同一物理原理:剪切波在坚硬或硬组织中传播更快,在柔软组织中传播更慢,因此测量波速即可得出以kPa表示的硬度值。9 在门静脉高压中,脾静脉压力升高引起脾脏淤血(脾大的淤血成分)和内脏高动力循环,从而增加脾脏硬度;这就是该测量的生理学基础。5 因此,SSM通过脾脏而非肝脏间接追踪门静脉压力,且即使在肝脏硬度未升高时也可升高:在乙型或丙型肝炎感染中,即使无纤维化,脾脏硬度的增加也可能早于肝脏硬度的增加。1 由于正常脾脏比正常肝脏更硬,瞬时弹性成像的设置需针对脾脏进行调整,包括改变剪切波频率、超声脉冲重复频率、测量深度和输出范围。7

操作方法

在所引用的VCTE方案中,患者测量前至少禁食4小时;该方案还规定脾脏测量时取仰卧位、左臂最大程度外展、肋间放置探头、屏气5秒,并至少重复测量10次,以kPa报告结果。10 对于点SWE,WFUMB建议同一部位获得5至10次有效测量,而EFSUMB建议至少10次;对于二维SWE,WFUMB建议从分离的视野中进行3至5次独立采集,实际操作中约在3至4个视野中各采集2至3次,共获得约10次有效测量。11 质量控制使用四分位距与中位数比值(IQR/中位数),以m/s表示的点SWE可接受≤15%,以kPa表示的二维SWE可接受≤30%。11 感兴趣区应位于器官包膜下1至2 cm至6 cm处,并应避开B模式图像上的大血管。12

起源

安东尼奥·科莱基亚及其同事在2012年发表于《Gastroenterology》的研究中报告了早期基于瞬时弹性成像的脾脏硬度测量,该研究使用TE评估HCV相关肝硬化患者的门静脉高压和食管静脉曲张,显示SSM与HVPG全范围值之间强相关;而塔尔沃卡及其同事已于2009年评估了在体MR弹性成像脾脏硬度测量。13 前期工作已用其他模态确立了可行性:贾扬特·A·塔尔沃卡及其同事于2009年在《American Journal of Roentgenology》上评估了在体MR弹性成像脾脏硬度测量对门静脉高压的诊断价值,14 广冈正及其同事于2011年在《Radiology》上报告了实时组织弹性成像测量的脾脏弹性作为门静脉高压标志物,15 宅间良孝及其同事于2012年在《Gastroenterology》上报告了声辐射力脉冲成像测量的脾脏硬度可识别伴有食管静脉曲张的肝硬化患者。16 科莱基亚及其同事后来提出了基于SSM和Baveno VI标准的联合模型,用于排除进展性慢性肝病中的高危静脉曲张(《Journal of Hepatology》,2018年)。17 塞西尔·巴斯塔尔及其同事于2018年报告了专用的脾脏硬度FibroScan检查,18 霍里亚·斯特凡内斯库及其同事于2019年在《Liver International》上表明该脾脏专用FibroScan改善了高危食管静脉曲张的筛查。19 这一时期的一个注意事项:早期TE-SSM研究仅纳入代偿期肝硬化患者。20

变体

瞬时弹性成像(VCTE,FibroScan)。 第一代脾脏VCTE使用50 Hz探头,硬度上限为75 kPa,可能高估脾脏硬度;专用的SSM@100 Hz模块工作范围为6至100 kPa,测量深度为25至55 mm。21 100 Hz探头的最佳截断值尚未确定,二维SWE和p-SWE测量SSM的可靠性标准也尚未建立。5

点剪切波和二维剪切波弹性成像。 SWE集成于高端超声设备中,使用B模式成像引导感兴趣区选择,并在组织深部产生波;分为点SWE和二维SWE。1

MR弹性成像。 在比较研究中,三维MRE脾脏硬度与HVPG相关性最强(r=0.686 r = 0.686 ,p<0.001 p < 0.001 ),CSPH诊断性能最佳(AUC = 0.911),优于二维MRE(AUC = 0.845)和SWE脾脏硬度(AUC = 0.583)。22

应用

Baveno VII推荐TE-SSM <21 kPa排除、>50 kPa确诊病毒性肝炎相关cACLD中的CSPH;这些阈值已被Baveno VIII进展性慢性肝病和门静脉高压共识会议的更新指南所取代,读者应参阅当前共识标准。1 WFUMB建议将SSM纳入联合算法:VCTE SSM <21 kPa可在同时具有LSM ≤15 kPa和血小板≥150 × 10⁹/L的患者中排除CSPH,SSM >40 kPa可在LSM ≥25 kPa和/或血小板<150 × 10⁹/L时确诊CSPH;SSM应与LSM联合评估和解读。5 2023年发表于《Lancet Gastroenterology & Hepatology》的个体患者数据荟萃分析(17项研究,1245例代偿期ACLD患者)在瞬时弹性成像队列(n=600 n = 600 )中验证了Baveno VII算法排除CSPH(NPV 100%,敏感性100%)和确诊CSPH(PPV 95%,特异性94%)的性能,但48%(95% CI 44–52)的患者落入不确定灰区;SSM双截断值模型将这一比例降至32%,单截断值40 kPa模型降至9%(敏感性93%,特异性86%,PPV 92%,NPV 85%)。6 一项纳入548例患者的随机试验使用LSM ≥12.5 kPa和/或SSM ≥41.3 kPa触发内镜检查,在检出任何静脉曲张(18.6% vs 24.5%)和需治疗静脉曲张(4.0% vs 5.8%)方面不劣于内镜优先筛查,平均41个月随访期间静脉曲张出血率相似。5 已发表的准确性估计存在差异:2018年一项荟萃分析报告诊断CSPH的准确性极佳(AUROC = 0.92),1 而Cochrane综述发现VCTE测量的SSM在固定特异性90%时敏感性为72.9%,在固定敏感性90%时特异性为80.6%,证据确定性极低。23 这种差异意味着临床医生应将SSM视为联合算法中的排除和确诊辅助工具,而非独立的诊断试验。

局限性与替代方案

可行性。 历史上,SSM失败率7%至24%是主要局限,脾脏专用100 Hz模块降低了失败率;24 其他综述报告失败率为13%至17%,在前后径小于4 cm的小脾脏中尤为困难。21 使用100 Hz探头时,257例患者中15例SSM失败(成功率94.16%),脾脏长度越短、体积越小、BMI越高,失败概率越高。25 在MASLD队列中,SSM失败率为19%,而LSM为5%,无效SSM与较高BMI、较大腰围和较低纤维化分期相关。26 由于瞬时弹性成像缺乏B模式成像,精确放置探头困难;一种结合TE与超声融合成像的3D打印装置将小脾脏(<100 mL)患者的成功率从35%提高至76.9%。27

混杂因素。 急性髓系白血病和骨髓纤维化等血液系统疾病会增加脾脏硬度。存在腹水时瞬时弹性成像结果准确性降低,急性肝炎、肝外胆汁淤积或进食后可出现假阳性;重测信度数据有限。8

晚期疾病。 在以失代偿期疾病为主的队列中,剪切波弹性成像对肝脏和脾脏的技术成功率均远优于TE(97% vs 44%和42%,p<0.001 p < 0.001 ),但两种方法测量的脾脏硬度均与CSPH无关;作者将这一差异归因于早期研究仅纳入代偿期肝硬化,而在较晚阶段硬度可能失去相关性。20 也有报告称,在24例门静脉高压较严重(HVPG ≥19 mmHg)的患者亚组中,脾脏硬度与HVPG不相关。20

替代方案。 健康成人SSM均值约为18.35 kPa,值<35.8 kPa可排除高危静脉曲张,而SSM >41.3 kPa与静脉曲张出血风险增加相关。8 与肝脏硬度、血小板计数、脾脏大小和内镜检查相比,SSM的定位是血流动力学辅助手段:HVPG测量仍是参考标准,但有创且可及性有限。4 一项比较LSM和SSM与HVPG的近期研究发现LSM的预测值优于SSM,SSM的可接受IQR变异度和所需测量次数尚需进一步验证。8

References

  1. Spleen stiffness measurement as a non-invasive assessment in patients with portal hypertension (BMJ eGastroenterology)
  2. Recommendations for portal hypertension assessment using elastography (Chinese Medical Journal)
  3. Refining portal hypertension assessment: The clinical significance of spleen stiffness measurement in the Baveno VII Era (Clinical and Molecular Hepatology)
  4. abstract (thelancet.com)
  5. WFUMB Guideline/Guidance on Liver Multiparametric Ultrasound: Part 1. Update to 2018 Guidelines on Liver Ultrasound Elastography
  6. Accuracy of spleen stiffness measurement for the diagnosis of clinically significant portal hypertension in patients with compensated advanced chronic liver disease: a systematic review and individual patient data meta-analysis (Lancet Gastroenterology & Hepatology)
  7. Outstanding feasibility of spleen stiffness measurement by 100-Hz vibration-controlled transient elastography
  8. Non-invasive Assessment of Clinically Significant Portal Hypertension (Hepatology International review, 2023)
  9. Liver MR Elastography Technique and Image Interpretation: Pearls and Pitfalls (RadioGraphics)
  10. Liver and spleen stiffness measurement using the new FibroScan module for prediction of esophageal varices and monitoring response to beta blockers in portal hypertension (Egyptian Journal of Radiology)
  11. Shear-Wave Elastography: Principles, Techniques, and Clinical Applications
  12. EFSUMB Guidelines and Recommendations on the Clinical Use of Liver Ultrasound Elastography, Update 2017 (Long Version)
  13. Antonio Colecchia and colleagues (2012). Measurement of Spleen Stiffness to Evaluate Portal Hypertension and the Presence of Esophageal Varices in Patients With HCV-Related Cirrhosis. Gastroenterology.
  14. Jayant A. Talwalkar and colleagues (2009). Feasibility of In Vivo MR Elastographic Splenic Stiffness Measurements in the Assessment of Portal Hypertension. American Journal of Roentgenology.
  15. Masashi Hirooka and colleagues (2011). Splenic Elasticity Measured with Real-time Tissue Elastography Is a Marker of Portal Hypertension. Radiology.
  16. Yoshitaka Takuma and colleagues (2012). Measurement of Spleen Stiffness by Acoustic Radiation Force Impulse Imaging Identifies Cirrhotic Patients With Esophageal Varices. Gastroenterology.
  17. Antonio Colecchia and colleagues (2018). A combined model based on spleen stiffness measurement and Baveno VI criteria to rule out high-risk varices in advanced chronic liver disease. Journal of Hepatology.
  18. Cécile Bastard and colleagues (2018). A Novel FibroScan Examination Dedicated to Spleen Stiffness Measurement. Ultrasound in Medicine & Biology.
  19. Horia Stefanescu and colleagues (2019). A novel spleen‐dedicated stiffness measurement by FibroScan® improves the screening of high‐risk oesophageal varices. Liver International.
  20. Prospective Comparison of Spleen and Liver Stiffness by Using Shear-Wave and Transient Elastography for Detection of Portal Hypertension in Cirrhosis (Elkrief et al., Radiology 2014)
  21. The future of splenic elastography: Is it a valuable tool for personalized treatment? (Revista Gastroenterología de México)
  22. MR elastography outperforms shear wave elastography for the diagnosis of clinically significant portal hypertension
  23. How accurate is liver and spleen stiffness measured by different elastography techniques in diagnosing clinically significant portal hypertension? (Cochrane)
  24. Advances in imaging, Elastography (Hepatology review)
  25. Enhancing liver cirrhosis varices and CSPH risk prediction with spleen stiffness measurement using 100-Hz probe (Scientific Reports)
  26. Role of Spleen Stiffness Measurement in the Evaluation of Metabolic Dysfunction-Associated Steatotic Liver Disease (Dig Dis Sci)
  27. Development of a method for measuring spleen stiffness by transient elastography using a new device and ultrasound-fusion method (PLOS One)

Initially written Sep 29, 2026 · Reviewed: Sep 30, 2026 · Edited: Sep 30, 2026; Oct 15, 2026; Oct 21, 2026

© 2026 Biostate AI。依据 Edgepedia Community License,注明出处即可免费使用。 嵌入参考卡片。

报告错误

脾脏硬度测量

请至少选择一个原因。